Technical data
Ceraphant PTC31B, PTP31B, PTP33B IO-Link
78
Hauser
Maximum turn down which can be ordered for absolute pressure and gauge pressure
sensors
Devices for gauge pressure measurement
• 6 bar (90 psi), 16 bar (240 psi), 25 bar (375 psi): TD 1:1 to TD 2.5:1
• All other measuring ranges: TD 1:1 to TD 5:1
Devices for absolute pressure measurement
• 100 mbar (1.5 psi), 250 mbar (4 psi), 400 mbar (6 psi): TD 1:1
• 1 bar (15 psi): TD 1:1 to TD 2.5:1
• All other measuring ranges: TD 1:1 to TD 5:1
Metal process isolating diaphragm
Sensor
Device
Maximum
Sensor measuring range
Lowest
calibratable
span
MWP
OPL
Factory settings
Option
lower (LRL) upper (URL)
[bar (psi)]
[bar (psi)]
[bar (psi)]
[bar (psi)]
[bar (psi)]
Devices for gauge pressure measurement
400 mbar (6 psi)
PTP31B
PTP33B
–0.4 (–6)
+0.4 (+6)
0.4 (6)
1 (15)
1.6 (24)
0 to 400 mbar (0 to 6 psi)
1F
1 bar (15 psi)
PTP31B
PTP33B
–1 (–15)
+1 (+15)
0.4 (6)
2.7 (40.5)
4 (60)
0 to 1 bar (0 to 15 psi)
1H
2 bar (30 psi)
PTP31B
PTP33B
–1 (–15)
+2 (+30)
0.4 (6)
6.7 (100.5)
10 (150)
0 to 2 bar (0 to 30 psi)
1K
4 bar (60 psi)
PTP31B
PTP33B
–1 (–15)
+4 (+60)
0.8 (12)
10.7 (160.5) 16 (240)
0 to 4 bar (0 to 60 psi)
1M
10 bar (150 psi)
PTP31B
PTP33B
–1 (–15)
+10 (+150)
2 (30)
25 (375)
40 (600)
0 to 10 bar (0 to 150 psi)
1P
40 bar (600 psi)
PTP31B
PTP33B
–1 (–15)
+40 (+600)
8 (120)
100 (1500)
160 (2400) 0 to 40 bar (0 to 600 psi)
1S
100 bar (1
500 psi)
PTP31B –1 (–15)
+100 (+1500) 20 (300)
100 (1500)
160 (2400) 0 to 100 bar (0 to 1
500 psi) 1U
400 bar (6
000 psi)
PTP31B –1 (–15)
+400 (+6000) 80 (1200)
400 (6000)
600 (9000) 0 to 400 bar (0 to 6
000 psi) 1W
Devices for absolute pressure measurement
400 mbar (6 psi)
PTP31B
PTP33B
0 (0)
0.4 (+6)
0.4 (6)
1 (15)
1.6 (24)
0 to 400 mbar (0 to 6 psi)
2F
1 bar (15 psi)
PTP31B
PTP33B
0 (0)
1 (+15)
0.4 (6)
2.7 (40.5)
4 (60)
0 to 1 bar (0 to 15 psi)
2H
2 bar (30 psi)
PTP31B
PTP33B
0 (0)
2 (+30)
0.4 (6)
6.7 (100.5)
10 (150)
0 to 2 bar (0 to 30 psi)
2K
4 bar (60 psi)
PTP31B
PTP33B
0 (0)
4 (+60)
0.8 (12)
10.7 (160.5) 16 (240)
0 to 4 bar (0 to 60 psi)
2M
10 bar (150 psi)
PTP31B
PTP33B
0 (0)
10 (+150)
2 (30)
25 (375)
40 (600)
0 to 10 bar (0 to 150 psi)
2P
40 bar (600 psi)
PTP31B
PTP33B
0 (0)
+40 (+600)
8 (120)
100 (1500)
160 (2400) 0 to 40 bar (0 to 600 psi)
2S
100 bar (1
500 psi)
PTP31B 0 (0)
+100 (+1500) 20 (300)
100 (1500)
160 (2400) 0 to 100 bar (0 to 1
500 psi) 2U
400 bar (6
000 psi)
PTP31B 0 (0)
+400 (+6000) 80 (1200)
400 (6000)
600 (9000) 0 to 400 bar (0 to 6
000 psi) 2W
1)
Highest turn down that can be set at the factory: 5:1. The turn down is preset and cannot be changed.
2)
Other measuring ranges (e.g. –1 to +5 bar (–15 to 75 psi)) can be ordered with customer-specific settings (see the Product Configurator, order
code for "Calibration; Unit" option "U"). It is possible to invert the output signal (LRV = 20 mA; URV = 4 mA). Prerequisite: URV < LRV
3)
Product Configurator, order code for "Sensor range"
4)
Vacuum resistance: 0.01 bar (0.145 psi) abs