Description of device parameters
Ceraphant PTC31B, PTP31B, PTP33B IO-Link
72
Hauser
Device Access Locks.Data Storage Lock
Activation/deactivation of DataStorage
1)
The "Device Access Locks.Data Storage Lock" parameter is an IO-Link standard parameter. The name of the parameter may be available in the
language configured in the IO-Link operating tool that is used. The display depends on the particular operating tool.
Navigation
IO-Link: Parameter → System → User Administration (ADM) → Device Access Locks.Data
Storage Lock
Description
The device supports DataStorage. If a device is being replaced, this allows the configuration
of the old device to be written to the new device. If, when a device is being replaced, the
original configuration of the new device is to be retained, the
Device Access Locks.Data
Storage Lock
parameter can be used to prevent the parameters from being overwritten. If
this parameter is set to "true", the new device does not adopt the data stored in the master's
DataStorage.
Options
• false
• true
LCK
unlocking code
Navigation
Display: EF → ADM → LCK
IO-Link: Parameter → System → User Administration (ADM) → LCK
Description
The device is unlocked by entering the code defined in COD.
Input range
0000-9999
Note
If parameters are locked, the word "LCK" appears on the local display as soon as an attempt
is made to change a parameter. Locking is enabled again after 60 seconds in the measured
value display and following a device restart.
COD
locking code
Navigation
Display: EF → ADM → COD
IO-Link: Parameter → System → User Administration (ADM) → COD
Description
A code can be entered to protect parameter settings against unauthorized and unwanted
access.
Input range
0000: Device is permanently unlocked
0001-9999: Device is locked
Factory setting
0000
Note
Locking is enabled after 60 seconds in the measured value display and following a device
restart.