System integration
Ceraphant PTC31B, PTP31B, PTP33B IO-Link
34
Hauser
8.2.2
IO-Link-specific device data
Designation
ISDU (dec)
ISDU (hex)
Size (byte)
Data type
Access
Default value
Serial number
21
0x0015
max. 16
String
r/-
Firmware version
23
0x0017
max. 64
String
r/-
ProductID
19
0x0013
max. 64
String
r/-
PTx3xB
ProductName
18
0x0012
max. 64
String
r/-
Ceraphant
ProductText
20
0x0014
max. 64
String
r/-
Absolute and gauge pressure
VendorName
16
0x0010
max. 64
String
r/-
Hauser
VendorId
7 to 8
0x0007 to 0x0008
r/-
17
VendorText
17
0x0011
max. 64
String
r/-
People for Process Automation
DeviceId
9 to 11
0x0009 to 0x000B
r/-
0x000700
Hardware revision
22
0x0016
max. 64
String
r/-
Application Specific Tag
24
0x0018
32
String
r/w
Actual Diagnostics (STA)
260
0x0104
4
String
r/-
Last Diagnostic (LST)
261
0x0105
4
String
r/-
8.2.3
System commands
Designation
ISDU (dec)
ISDU (hex)
Value range
Access
Reset to factory settings (RES)
130
0x0082
w
Device Access Locks.Data Storage Lock
12
0x000C
0 ~ False
2 ~ True
rw
Device Access Locks.Local Parametrization Lock
130
w
8.3
Overview of diagnostic events