Sampling modes:
6 individual presettable programmes, presettable programme change
(switching) conditions (eg. Q-t changeover, Q-q changeover, etc.)
Possibilities:
Time proportional
Quantity proportional
Event controlled
Manual start
Timer:
Individual start/stop operation using daily/weekday switch functions
Sample distribution:
Via presettable filling time, number of samples in bottle or container or
external Signal.
With presettable sequence end after last bottle or continuous operation.
Impulse input:
Optocoupler input: Positively flanked, galvanically isolated,
min. impulse length 10 ms
Low: 0 ... 3 Volt
High: 12 ... 30 Volt
Analogue input:
Switchable as current or voltage input:
Current input max. load 50 Ohm:
0 ... +20mA
4 ... +20mA
Voltage input load 1 Megaohm:
0 ... +1 Volt
(Select via switch)
0 ... +10 Volt
Stop input:
Optocoupler input: Galvanically isolated, stop when high
Low: 0 ... 3 Volt
High: 12 ... 30 Volt
also used for distribution tap changeover.
Control input:
Optocoupler input: Galvanically isolated. Can be set up as programme
change or event input.
Programme change on High.
Programme return on Low.
Event active on positive edge.
Low: 0 ... 3 Volt
High: 12 ... 30 Volt
Three outputs:
Output 1, 2 and 3: Transistor outputs NPN
Switch condition dependent on settings
"Standard" or "Inverse"
Standard:
Power on, active (alarm condition)= Off
Power on, not active (no alarm)= On
Power off
= Off
Inverse:
Power on, active (alarm condition)= On
Power on, not active (no alarm)= Off
Power off
= Off
Interface:
TTY:
Formated for
Primo-Bit data printer
V24 (RS232):
Option
Auxiliary voltage from unit:
U
ext
: + 20 bis + 30 VDC (200 mA)
Technical data
asp-port d 2 se
40