Base settings
110
Sample
volume
Set up
range:
0...500 ml
111
Bottle
volume
Set up
range:
00,0...99,9 l
112
Cleaning
times
Empty:
1...120s
Rinse:
1...120s
113
Empty
cycle start
0...99
114
Container
Cleaning
No.
# 1
# 2
115
Door open
counter
6 digits
resettable
120
Self
optimising
sample
phases
ON / OFF
121
Blow out time
(Phase 1)
Set up
range:
0...15 sec.
122
Suction time
(Phase 2)
Set up
range:
10...360sec
123
Dosing time
(Phase 3)
Set up
range:
5...150 sec.
124
Sample delay
Set up
range:
0...100 sec.
125
Conductivity
sensitivity
low
middle
high
126
Synchronisation
AUT / time
127
Synchronisation
time
Hours
00 .. 24
Minutes
00 ... 60
130
Thermostat
On / Off
131
Preset
temperature
0...30 °Cels.
132
Automatic
defrost
Yes / No
132
Defrost time
Range
1 ... 999 min.
Cycle range
2 ... 999 min.
140
Select signal
input type
Analogue
Impulse
None
141
Select
analogue input
0...20mA
4...20mA
0...1 V
0...10V
142
Set analogue
input values
I/Sec or m
3
/h
Range:
1...9999
143
Set analogue
input value
decimal point
Without, after
1. to 3. points
144
Set impulse
input values
I/Impulse or
m
3
/Impulse
1...9999
145
Set impulse
input values
decimal point
Without, after
1. to 3. points
150
Set up
output 1
One from
eight pos.
see description
151
Set up
output 2
One from
eight pos.
see description
152
Set up
output 3
One from
eight pos.
see description
160
Select
interface
Primo-Bit
or
Option
161
Print sample
report
On or Off
162
Print
parameter
report on start
On or Off
163
Installation
name (unit
identifier)
10 digit
selectable
164
Serial unit
address
Selectable:
00...99
165
Select
interface type
TTY / (V24)
166
Set Baudrate
Selection
300 / 600 /
1200 / 2400
167
Select parity
Selection
odd /even /
mark / space
168
Select Stop-Bit
Selection
1 / 2
169
Display of
number of
data Bits
7
170
Date
2 digits each
Day, month,
year
171
Time
Hours
00...24
minutes
00...60
172
Summer/
normal time
changeover
automatic/
manual/
switched off
173
Advance (1h)
NT
→
ST
Selection:
Day, month,
year; hour,
minute
174
Reverse (1h)
ST
→
NT
Selection:
Day, month,
year; hour,
minute
Addr.
Description
Works settings
110
Set up the same dosing volume as on the filling tube in the dosing chamber
200 ml
111
Set up single bottle / container volume (-10% for safety) (Addr. 110/111 operate as bottle overfill security).
10,0 l
112
Set up the times in the various cleaning cycles.
Empty: time for emptying the container
Rinse: time during which water is used
30 s
30 s
113
The number of samplers before container change when the cleaning and drain cycle of the new container starts.
Important when distribution operates on number of samples per container.
Presettable from 0 - 99
2
Operation and display
asp-port d 2 se
26