Cabinet:
Hard polyeurythane foam
H x W x D approx. 725 x 532 x 400 mm
Weight:
Approx. 30 kg
Protection class:
Controller (Keypad): IP 55 to DIN 40050
Allowable ambient temperature:
Without heater
+0°C ... +40
o
C
With heater
-15°C ... +40°C
Allowable medium temperature:
>0
o
C ... +50
o
C
Minimum liquid conductivity:
≥
30
µ
S/cm (others optional)
Power supply:
230 V AC + 10% -15%, 50/60 Hz or
Power consumption:
AC version:
Without heater 50 W
Safety:
To VDE 0411 Part 1 protection class I, overvoltage categorie II
EMC/immunity:
To EN 50082-1
RF:
To EN 55011, class A (Industrial surroundings)
Data security:
Up to 500h during power failure
(condition: powered for at least 7 days previously)
Security code number:
All entries and data protected against unauthorised tampering by means of a
security code number. The code number to unlock the programme for setting
up is "6051"
Feed system:
Built in diaphragm pump
Feed conditions:
Feed height:
max. 6m
at 1013 hPa
Feed distance:
max. 30m
at 1013 hPa
Suction velocity:
0.6m/sec,
13 mm diameter hose
Sample volume:
20 ml to 200 ml presettable (option 350ml)
asp-port d 2 se
Technical data
39