Fault messages
The sampler contains a self monitoring function system. Faults occurring are
displayed including a fault number as well as a hint as to how to remove the fault.
Once the fault has been removed acknowledge this by operating the ON push button.
Fault-
No. #
Display
Cause and cure
01 Composite fault
Follow all steps as in # 03 to # 09 in
See operating instructions
order to cure.
If no cure: contact E+H Service.
03 Electrode 1 short circuit
Remove dosing chamber, flange and
Clean dosing system
electrodes (underside of flange)
clean thoroughly.
04 Electrode 2 active
As # 03 Careful: = Safety switch,
Clean electrodes
must be cleaned !
The manufacturer does not accept any liability for any damage
(flooding on sampling) occurring to the unit if this fault has been
acknowledged without first cleaning the dosing system and electrodes !
05 Air manager
Edge connector loose or pneumatic
see operating instructions
controller defective.
06 Distribution tap zero point
Distribution system defective,
Exchange distribution system or have
unit repaired by E+H Service.
07 Distribution tap manipulated
Distribution tap mechanically blocked
see operating instructions
or manually moved (>7,5
o
)
Distribution tap is self-positioning.
08 Distribution tap connector
Connection from distribution system to
Check connection
controller must be solid.
09 Input current <3mA
Cable open circuit on 4..20 mA.
Check connection
Check cable and meter.
10 Battery low voltage
Only on 12VDC version.
Charge battery
11 Temperature
Temperature sensor not plugged in
Check sensor
or defective.
System error
Preset done
RAM data loss.
fatal error, E+H Service.
Reset done
RAM data loss.
Unit too long without power;
Set up user data new.
Internal accumulator (empty or),
defective, E+H Service.
Calibration missing
Recalibrate analogue or temperature
input, E+H Service.
Unit fault
Check power supply and that the
ambient temperature range is not
exceeded.
If fault reoccurs:
E+H Service.
Faults # 03 and # 04 are meant as maintenance displays.
The four digit fault number, in brackets, is for the decoding of multiple and system
errors by E+H service personnel.
asp-port d 2 se