Description
410
Date of last service. Must be set to "Yes" and initiated by sevice technician !
411
Please always indicate this value on any queries !
412
Sum of all faults that led to an abort. Last fault is indicated with an error code number.
420
Total unit running time (time connected to mains power).
421
Pump running time (pump running time
÷
unit running time = use ratio)
This counter should be reset to zero by the service technician when changing the pump.
422
Cooler running time (cooler running time
÷
unit running time = use ratio)
This counter should be reset to zero by the service technician when changing the cooler.
423
Number of sample cycles.
424
Number of times the safety electrode (electrode 2) switched the unit off.
425
Number of times this was acknowledged without cleaning the electrode.
(Note: We the manufacturer cannot accept any liability, this includes
guarantee claims for any damage occurring).
430
Test phases: Start, go to zero point, close hose clamp, blow out, suck, dose, open hose clamp,
zero point.
431
Display of actual distribution tap position. Change tap position by operating the ENTER push button.
440
bis
443
Calibrate analogue input see matrix.
450
bis
453
Calibrate temperature input see matrix.
460
bis
465
Unit test (quick check) see matrix.
470
RESET means:
All dynamic data (eg. counter values) are reset.
471
PRESET means:
All programme settings, data and counter values are deleted
and reset to works settings. Therefore BE SURE !
asp-port d 2 se
Operation and display
33