Address
Description
Works settings
310
Select one from six possibilities:
Continuous operation:
Once the AUT push button has been operated the unit
operates continuously with the preset programme (addresses 320 - 369 not active).
No start / stop function !
Once:
1 start and stop time including date.
(set up in addresses 320/321 and 330/331).
Daily:
A maximum of 5 start and stop times can be set up
and are valid for each day (addresses 340 to 349).
Daily with date:
Each can have 5 start and stop times set up and is
valid daily between the preset start and stop dates.
(Set up in addresses 320/321 and 340/349).
Week day:
A maximum of 10 independent start and stop times can be
set up. These include weekdays (Monday to Sunday).
This data is set up in addresses 350 to 369.
This function is repeated weekly.
Week day with date:
Set up start date (addr.320) and stop date (addr.321).
Now set up a maximum of 10 independent start and stop times
Each with a weekday, ie. Monday to Sunday (addr.350 to 369).
The sampler operates from the start date to the stop date.
Within these two dates it operates using the preset start and stop times.
Note: The start and stop times are repeated weekly if the start and
stop dates are far enough apart.
Continuous
311
Yes:
All start/stop times are reset to zero.
No:
The start/stop times remain unchanged.
-
320
bis
369
See address 310
Not active
General information to the
start/stop operation mode.
The start stop operation mode is linked to the automatic mode (operation of
the "AUT" push button). Date/time as well as the heating and cooling cycles
continue to operate.
The automatic sequence starts or restarts on reaching the start time (or
continues to operate). The times of sample cycle start and distribution bottle
change, if set to timed, are always referred to the start time.
For example:
Start time Monday 00:00, time cycle 15 min.
Distribution 2 hours.
This means that:
Sampling occurs on Monday at 00:15, 00:30, 00:45 etc.,
Distribution at 02:00, 04:00, 06:00 etc..
The automatic sequence is stopped once the stop time is reached.
On the next start time (at timed sampling and distribution) the sample cycle
timer (eg. 15 min.) restarts at zero, the distribution timer continues to operate
as normal (from distribution time at stop time).
The conditions set for the "Time proportional sampling" are valid for the
sample cycle counter when set to "Quantity proportional sampling". Also
the same conditions are valid for the distribution counter whether set to
"Timed bottle change" or "Number of samples bottle change".
Recommendation
Combine the stop time with the distribution switch times.
asp-port d 2 se
Operation and display
31