Troubleshooting
PlasmaQuant 9100 (Elite)
54
¡
Viscous sample solution/higher den-
sity than calibration solution
¡
Matrix adjustment (add to calibra-
tion solutions or dilute)
¡
Use one/several internal standards
¡
Spray chamber filled
¡
Empty spray chamber
¡
Check drainage pump hose, select
larger diameter if necessary
Measured value too high
Cause
Remedy
¡
Calibration fault
¡
Check calibration solution
¡
Peak position slightly shifted or non-
peak measurement
¡
Spectral interference overlooked.
Use another analytical line or acti-
vate fault correction
¡
Contamination/carry-over
¡
Find and remedy cause for contami-
nation/carry-over
¡
Concentrations appear higher due to
volatile substances
¡
Optimize sample preparation
¡
Analyte is an alkaline metal (or an
easily excitable atomic line)
¡
Alkaline effect. Optimize plasma
temperature (nebulizer gas flow
and/or performance) and plasma
observation
¡
Fault when using an internal stan-
dard
¡
Wrong standard dosage
¡
Insufficient reaction adjustment to
plasma temperature change. Matrix
adjustment and better fit between
the analytical line behavior and the
internal standard
¡
Warm-up phase not observed
¡
Wait for warm-up phase before cali-
bration
¡
Sample foams when shaken
Surface active substances in the mea-
surement solutions:
¡
Optimize sample preparation
¡
Add surface active substances to cal-
ibration solutions as well
Poor precision
Cause
Remedy
¡
Pump fast mode was active shortly
before the measurement
¡
Limit fast mode to the time neces-
sary to transfer measurement solu-
tion to nebulizer
¡
Pre-purge time too short
¡
Extend pre-purge time
¡
Nebulizer or injector clogged
Check with Na solution (1 g/L), if neces-
sary. If no plasma coloration (orange) is
observed:
¡
Check that nebulizer nozzles are un-
clogged and clean nebulizer
¡
If sample nozzle is clogged, filter so-
lutions or use inline filter
¡
If argon nozzle is clogged, dilute
measuring solutions or use argon
humidifier
¡
Nebulizer gas flow not optimal
¡
Optimize nebulizer gas flow