PlasmaQuant 9100 (Elite)
Installation and commissioning
35
Installing the sampler
NOTICE
Risk of damage to the sensitive electronics
¡
Only connect the sampler to the electrical grid after installation.
}
Place the tray on the automatic sampler foot, and place the base plate for
holding the sample racks on that.
In this case, the place for the purging vessel must be located on the left
to the rear. The base plate is correctly mounted if it fails to move after
light shaking.
}
Installing the purge vessel: Insert the purging vessel into the rear left re-
cess and turn it clockwise by 90°.
}
Fit the racks for special samples (1) on the base plate and attach the re-
quired sample racks (2).
The sample positions are coded with a three-digit number (e.g. 108) in
the control software. The first number indicates the sample rack itself,
the following numbers the position on the rack. The first sample rack is
located in front of the purging vessel, followed by the second and third.
In the software the positions are only shown for illustration.