Installation and commissioning
PlasmaQuant 9100 (Elite)
44
4.4.4
Installing the inline filter
1
2
3
Figure 28Installing the inline filter
1 Sample hose to the nebulizer
2 Filter block with hose connectors
3 Sample pump hose
}
Insert the inline filter in the hoses so that the arrow on the filter block point in the
flow direction (i.e. in the direction of the nebulizer).
}
Assemble the inline filter as described in the accompanying datasheet.
}
To do so, insert the nebulizer hose in the banjo bolt via a ferrule. The conical side of
the ferrule must point toward the banjo bolt.
}
Screw the hose connector into the outlet of the filter block.
}
Screw a short capillary hose into the inlet of filter block via hose connector.
}
Connect the capillary hose with the sample pump hose. To do this, push the capillary
hose into the pump hose.
}
Connect the nebulizer hose with the nebulizer.
Figure 29Fit of the ferrule in the banjo bolt
1 Ferrule
2 Banjo bolt
3 Hose