Installation and commissioning
PlasmaQuant MS product family
46
4.3.3
Installation of the spray chamber
1.
Attach a ~10 cm piece of drain tubing to
the waste outlet at the bottom of the
spray chamber.
2.
Open the holder assembly. Place the spray
chamber into position in the lower half of
the holder assembly.
3.
Place the top half of the holder assembly
in position over the spray chamber and
secure it using the two knurled knobs.
If the glass transfer tube is already in place:
4.
Swing the mounting arm inward towards
the transfer tube.
5.
Fit the spherical joint of the transfer tube
into the joint of the spray chamber and
secure it using the glassware clamp.
If the transfer tube is not in place, refer to the
installation instructions on the next page.