Installation and commissioning
PlasmaQuant MS product family
32
4.1.2
Power supply
WARNING
Shock hazard!
During electrical installation, observe the VDE (German Association for Electrical
Engineers) electrotechnical guidelines and local regulation requirements!
The mains supply must be correctly earthed. Do not use an adapter in the mains
cabling.
The PlasmaQuant MS is available as a single-phase electrical configuration. If a single-
phase electricity supply is not available, most three-phase supply systems can be
adapted to supply the correct line-to-neutral voltage in the specified range.
To enable the PlasmaQuant MS to be used in areas with power supply voltages of
100 V/115 V/120 V/127 V, connection to 2 phases is possible via optional accessories.
Contact Analytik Jena if required. This installation may only be carried out by service
engineers from Analytik Jena or by specialist personnel authorized and trained by
Analytik Jena.
Electrical power is connected from your supply outlet to the PlasmaQuant MS by a
single three meter 1 ph power cable (length 3 m, Ø min. 4 mm
2
Cu). With the
electrical power cable connected to your power outlet and the power turned on at the
power outlet, power can still be isolated at the instrument itself.
The PlasmaQuant MS incorporates a circuit breaker (located at the rear of the
instrument) and a standby switch (located on the front panel of the instrument) to
control power flow between the power outlet and the power distribution module of
the instrument. Both of these isolation devices must be in their ON position before the
instrument can function.
To avoid sudden voltage fluctuations, do not connect the PlasmaQuant MS to the
same electrical circuits as other power-intensive devices.
Summary 4
Power supply
Voltage
200-240 V AC ± 5 %
Frequency
50/60 Hz
Typical average power consumption
2700 VA
Max. power consumption
3700 VA
Line current
18 A max.
Fuse provided (mains side)
25 A, Curve C
Switching-on conditions