PlasmaQuant MS product family
Autosampler ASPQ 3300
121
The autosampler is installed next to the PlasmaQuant MS. Between these two
instruments 30 cm of clearance is required to allow space for opening the interface
door.
NOTICE
The autosampler must not be connected to the mains during installation.
1.
Place the tray on the autosampler foot and on it the base
plate for holding the sample racks.
In this case, the place for the wash cup must be located at the
rear on the left. The base plate is correctly mounted if it fails
to move after light shaking.
2.
Install the wash cup. Insert the wash cup into the rear left
recess and turn it by 90° in a clockwise direction.
3.
Fit the racks for standards and QC solutions (1) in the base
plate and attach the required sample racks (2).
Installing the
autosampler