Fault removal
multi N/C pharma UV
90
116
Unknown sampler command
Cause
Remedy
Communication error
Contact Service
201
Restart the internal program
Cause
Remedy
overvoltage
short-term power failure
initialize analyzer
401
402
403
404
Syringe pump initialization
Syringe pump: invalid command
Syringe pump: invalid operand
Syringe pump faulty command sequence
Cause
Remedy
communication error
initialize analyzer
syringe pump faulty
inform Service
407
Syringe pump: syringe pump not initialized
Cause
Remedy
analyzer not yet initialized after switching on
reset the syringe pump
initialize analyzer
409
Syringe pump: pump sluggish
Cause
Remedy
clogging of a hose line (6) or (7)
search for cause and remedy fault
replace hose line - if necessary remove
hose line and rinse with ultrapure water,
then reinstall
initialize analyzer
syringe pump faulty
inform Service
410
Syringe pump: valve sluggish
Cause
Remedy
syringe pump faulty
valve broken
inform Service
411
415
Syringe pump: pump step not permitted
Syringe pump: invalid command
Cause
Remedy
communication error
initialize analyzer
syringe pump faulty
inform Service
MESSx
Analyzer error: MESSx measurement is cancelled
Cause
Remedy
equipment fault
initialize analyzer
check window
System state
after error
removal
search for equipment fault and remove
error