multi N/C pharma UV
Operation
61
Fig. 32 Disabling individual measured values of a calibration
Enabling/disabling measured values for preparation water
The individual values determined for the preparation water can be viewed by
clicking on the button [E
DIT
]
and enabled/disabled for the calibration.
Adding measuring points
An existing calibration can be extended by additional measuring points. Carry out a
measurement with the same method (select C
ALIBRATION
as sample type and enter
the target concentration) and select the corresponding analysis report via the
button [A
DD MEASURING POINT
].
Measuring points can only be added individually.
6.4.4
Transferring calibration parameters to a method
Transfer calibration parameters to a method as follows:
1.
Select an appropriate calibration range for the respective parameters (e.g.
NPOC/TN).
Up to three calibration ranges for each parameter can be stored in a method. It
must be noted that the ranges should overlap and have no gaps! Every calibration
range must be transferred individually, i.e. select calibration range, link, and
transfer.
Using calibration function with quadratic regression only one calibration range can
be stored in a method.
2.
For each selected calibration range and analysis parameter to be transferred
enable the field
Use calibration
with (
).
Not all calibrated parameters need to be transferred to the method.
3.
Click on the button [L
INK WITH METHOD
].