multi N/C pharma UV
Connecting add-on devices
43
1 grounding conductor
2 switching power supply connection
3 RS 232 interface - analyzer connection
Fig. 20 Connections at the bottom of the sampler AS 21
1.
Screw the holder to the right-hand side of the analyzer.
2.
Place drive onto the holder.
3.
Plug the grounding conductor into the connection on the rear of the analyzer (8 in
Fig. 13 p. 23).
4.
Plug cable on the low voltage side of the table power supply included in the
delivery into the connection at the bottom of the sampler. Do not yet connect the
power supply to the mains.
5.
Plug the interface cable into the bottom of the equipment. Connect the sampler to
the interface at the rear of the analyzer (11 in Fig. 13 p. 23).
6.
Place a sample tube into position 1 of the sample tray.
7.
Insert the cannulas into the autosampler arm according to Fig. 19.
8.
Manually adjust the height of the cannulas so that the cannula tips protrude 1 –
2 cm over the tube edge and cannot touch the tubes when the autosampler arm is
moving. Secure the cannulas by slightly tightening the screw.
Attention: the screws must not bend the cannulas under any circumstances.
9.
Connect the switching power supply to the mains. Switch on the switching power
supply.
10.
Check the configuration via the menu command I
NSTRUMENT
SYSTEM INFORMATION
in the window
SET
-
UP INFO
. If necessary, modify the configuration:
Exit the program multiWin.
On the Windows user interface start the
SET
-
UP TOOL
under S
TART
P
ROGRAM
F
ILES
MULTI
W
IN
MULTI
W
IN SET
-
UP TOOL
.
In the list S
AMPLER
select the sampler type.
Exit the
SET
-
UP TOOL
with [C
REATE
].
Start the multiWin program and select the C
ONFIGURATION
E
DIT OPTIONS
menu
command to open the O
PTIONS
window on the
A
NALYZER COMPONENTS
tab. In the
S
AMPLER
group, select the correct tray and tube size. Exit the O
PTIONS
window
with [OK].
11.
Adjust the immersion depth of the cannulas in the tube (z direction) in multiWin.
Select the I
NSTRUMENT
S
AMPLER
A
LIGNMENT
menu command to open the
window with the same name.
Installation at the
analyzer