multi N/C pharma UV
Connecting add-on devices
41
12.
Adjust the cannulas so that approx. one third of the cannulas is visible above the
sleeve. Secure the cannulas by slightly tightening the screw.
13.
Connect the two connection hoses to the analyzer to the cannulas:
Hose no. 6 = sample aspiration hose
Hose no. 10 = purging hose for NPOC measurements
Release the upper Fingertight connection of the cannula.
Guide the hose through the banjo bolt.
Slide the conical nipple with the conical side towards the banjo bolt onto the
hose. The conical nipple and hose must be flush.
Retighten the Fingertight connection.
1 Conical nipple
2 Banjo bolt
3 Hose
Fig. 18 Hose in Fingertight connection
14.
Attach the sample cover (if present) so that it is positioned in the guide rail.
15.
Connect the power supply to the grid.
16.
Check the configuration in the multiWin program via the I
NSTRUMENT
S
YSTEM
INFORMATION
menu command in the
SET
-
UP INFO
window. If necessary, modify the
configuration:
Exit multiWin.
In the Windows user interface, start the
SET
-
UP TOOL
under S
TART
P
ROGRAM
F
ILES
MULTI
W
IN
MULTI
W
IN SET
-
UP TOOL
.
Select the sampler type in the S
AMPLER
list.
Exit the
SET
-
UP TOOL
with [C
REATE
].
Start the multiWin program and select the C
ONFIGURATION
E
DIT OPTIONS
menu
command to open the O
PTIONS
window on the
A
NALYZER COMPONENTS
tab. In the
S
AMPLER
group, select the correct tray and tube size. Exit the O
PTIONS
window
with [OK].
Before the first start the sampler must be adjusted (see "Adjusting the AS vario
autosampler" section, p. 69).
5.2
Sampler AS 21
WARNING
Make sure that no liquid reaches the cable connections or the inside of the equipment
or the power supply. There is a danger of electric shock.