multi N/C pharma UV
First commissioning
37
4.1.3
Energy supply
WARNING
The analyzer multi N/C pharma UV must only be connected to a properly grounded
mains outlet in accordance with the voltage specifications on the type plate!
The multi N/C pharma UV is operated from the single phase alternating current mains.
The installation of the electrical equipment of the laboratory must comply with the
standard DIN VDE 0100. After the connection point an electrical current in accordance
with the standard IEC 38 must be available.
4.1.4
Gas supply
The operator is responsible for the gas supply and the corresponding connections and
pressure reducers.
The connection hose with outer diameter 6 mm and inner diameter 4 mm is included
with the delivery.
4.2
Unpacking and placing the analyzer
CAUTION
The analyzer multi N/C pharma UV must only be set up, assembled and installed by
the customer service department of Analytik Jena GmbH or trained specialist personnel
authorized by Analytik Jena GmbH!
Any unauthorized intervention in the analyzer can endanger the user and the
operational safety of the equipment and limits or completely invalidates any warranty
claims.
ATTENTION
Retain the transport packaging! Return transport for service must be in the original
packaging. This alone prevents transport damage.
The analyzer multi N/C pharma UV is unpacked and assembled by the customer service
department of Analytik or its authorized and trained specialist personnel.
Please check when unpacking the device for completeness and soundness of the
delivery in accordance with the packing list included.
After assembly the customer service tests the analyzer and documents the test.