First commissioning
multi N/C pharma UV
36
4
First commissioning
4.1
Site requirements
4.1.1
Installation conditions
The following requirements are placed on the climatic conditions in the operating
room of the analyzer:
Temperature range: +10 °C to +35 °C
max. humidity:
90 % at 30 °C
Air pressure:
0.7 bar to 1.06 bar
The laboratory atmosphere should be as low as possible in TOC, nitrogen oxide and
dust and free of draft, corrosive vapors and vibration. Smoking is prohibited in the
operating room of the analyzer!
The following requirements are placed on the location of the analyzer:
Do not locate the analyzer directly near a door or window.
Place the analyzer onto a acid-proof surface.
Do not locate the analyzer near sources of electromagnetic interference.
Avoid direct sunlight and radiation from heaters onto the analyzer; if necessary
ensure air-conditioning.
Never obstruct the front doors, the left side wall and the ventilation slots of the
analyzer with other equipment or furnishings!
Keep a safety distance of at least 5 cm from the back and the right side of the
equipment to other equipment or walls!
4.1.2
Space requirement
The space required is a function of all components needed for the measurement.
Leave adequate space for the PC, monitor, printer and any add-on equipment.
The AS vario autosampler and the EPA sampler must be positioned on the right side of
the analyzer. The AS 21 and AS 10 samplers are attached to the right-hand side of the
analyzer. The layout of the other components can be adapted to the local conditions.