Interferential 960/Multidyne 970
31
Interrupted galvanic set-up
(Multidyne 970 only)
Treatment time:
is set as described before.
Form:
Three different pulse shapes are available: Rectangular), Triangular
and Trapezoidal. A full description of these waveforms is available in the
technical specification section of this manual.
Width:
The pulse width may be set from 1ms to 1s for all waveforms with
additional narrower pulses for rectangular only.
Rate:
Pulse rates of 2, 5, 10, 20 and 30 per minute are available.
Polarity:
Positive going, negative going or auto-reverse polarities are
available (auto-reverse occurs half way through the selected treatment
time).
When running, the output state is represented by a graphic in the top right
corner of the display -
CC/CV:
The output may be set to be constant current (CC) or constant
voltage (CV).