Broadband noise
>95
Temporal reflex characteristics
Initial latency: 20 ms ±15 ms
Rise and fall time: 30 ms ±15 ms
Terminal latency: 10 ms ±15 ms
Undershoot and overshoot: <10 %
Pulsed stimulus characteristics
Rise and Fall time: 5 ms
On time: 55 ms
Off time: 60 ms
Accuracy: 0 ms
Electromagnetic Compatibility (EMC)
Medical electrical equipment needs special precautions regarding EMC and needs to be installed
and put into service according to the EMC information provided in this section. Portable and mobile
radio frequency (RF) communications equipment can affect medical electrical equipment. Following the
guidelines in this section will help prevent this.
Warning:
The Kuduwave has been tested to the BS EN 60601-1-2:2015 for both immunity
(susceptibility to interference from external sources) and emissions (interference generated by the
Kuduwave). In order to ensure correct operation the following precautions must be adhered to:
The use of components and cables other than those specified or sold by eMoyo may result in increased
emission or decreased immunity of the Kuduwave. The list of cables and components below must be
adhered to in order to ensure compliance.
The Kuduwave should not be used adjacent to or stacked with other equipment. If adjacent or stacked
use is necessary then the Kuduwave™ should be observed to verify normal operation in the
configuration in which it will be used.
List of cables and components that affect compliance
●
USB Cable, Type A to mini-B, maximum length 3.0 meters.
Guidance and Manufacturer’s Declaration - Electromagnetic Emissions
The Kuduwave™ is intended for use in the electromagnetic environment specified below.
The customer or operator of the Kuduwave™ must ensure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The Kuduwave™ uses RF energy only for its internal
function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby
electronic equipment.
RF emissions CISPR 11
Class A
The Kuduwave™ is suitable for use in all establishments
other than domestic and those directly connected to
the public low-voltage power supply network which
supplies buildings used for domestic purposes
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations
/ flicker emissions
Not applicable
Page 90