respectively.
To the left and right of the test buttons are indicator lights.
A purple light indicates that a sweep is underway.
An orange light indicates that audio is being played but the sweep is not underway yet.
A white light indicates that the pump is initializing or resetting itself.
The pressure in the ear canal is updated in real time and is displayed on the pressure bars on each side.
The tympanogram is updated in real-time and is displayed on the graph in the center.
The blue line is for the left ear tympanogram. The red line is for the right ear tympanogram.
The ear canal volume is indicated as a
metric and is visually displayed on the bars
to the right of the graph (left bar for left ear,
right bar for right ear). The scale runs from
0 cm³ to 5 cm³.
If the seal is lost during a test, the sweep is aborted and the pump is reset. To delete a tympanogram
click on the Discard button
View the Tympanograms Metrics
Various metrics are displayed for the tympanogram for each side:
Page 57