Technical Specifications
Note
: These specs are for Kuduwaves manufactured after March 2022, from Serial # 0901-04511
onwards.
WARNING
:
No modification of this equipment is allowed.
Standards
The Kuduwave has been independently examined, tested and certified by a registered Notified Body in order to
ensure Safety and Design Standards detailed in the General and Audiometry Specifications.
Audiometry Standards
Pure tone: ANSI S3.6 (Puretone Type 2), BS EN 60645-1 (Type 3)
Tympanometry ANSI S3.39 (Type 1), BS EN 60645-5 (Type 1)
Other Standards
ANSI S3.21, ANSA S3.55, ANSI S3.13, BS EN 60601-1, BS EN 60601-1-2, BS EN
60601-1-6, BS EN ISO 13485, BS EN ISO 14971, BS EN 62304, BS EN ISO 14155,
BS EN ISO 15223-1, EN 1041
FDA Establishment No.
3011545613
FDA Owner/Operator No.
10048069
Product Code:
Device Class:
Regulation Number:
EWO
2
874.1020
Calibration
Laboratory calibrated in accordance with:
BS EN 60645-1, EN 60645-2, SANS 10154-1 and
SANS 10154-2
Instrument Specifications
Dimensions
210 x 260 x 110 mm
Shipping
dimensions
410 x 320 x 190 mm
Net weight
697 g
813 g (including Tympanometer functionality)
Shipping weight
1759 g
1930 g (including Tympanometer functionality)
Power supply
2x standard laptop USB ports (5 V, 900 mA max per port)
Unplug laptop from mains while testing
Data transfer
Twisted Kuduwave™ dual USB cable
2x standard 3 meter A Male to Mini B Male USB cables
Environmental
conditions
Indoor use only
Operational temperature
15 - 35 °C
Operational humidity (non-condensing)
30 - 90 %
Operational ambient pressure
98 - 104 kPa
Shipping and storage temperature
10 - 40 °C