6. Place and secure the source back cover by screwing the cover and securing the back
screw bolt.
4.10.2
Wiring to source terminals
The source contains six wiring terminals.
Table 4-6: Flash source wiring terminals
Terminal number
Function
1
Power + 24 Vdc
2
Return - 24 Vdc
3
Not used
4
Not used
5
RS-485 (+)
6
RS-485 (-)
4.11
Align detector
Use the alignment tool to perform full alignment.
Align the detector in two stages: crude alignment and fine alignment.
The alignment tool includes a periscope that consists of a prism and an ocular located
vertical to the alignment tool assembly. This allows you to look into the opposite unit
perpendicularly to the alignment when access from the rear of the unit is impossible. For
installations where rear access is possible, you don't need to install the periscope. In this
case, you can remove it by releasing the periscope fastening screw.
NOTICE
Prior to installing the alignment tool, verify that the alignment tool and its sight mounting
are free from any dirt to ensure proper alignment according to factory calibration.
Do not attempt to change any factory calibration at the alignment tool or its mounting.
To align the detector (see
1. Make sure that the detector and flash source are installed properly.
provides installation instructions.
2. Remove the front shield using the two captive screws.
3. Install the alignment tool assembly on the detector/source front.
4. Fasten the alignment tool with fastening screws.
4.11.1
Perform crude alignment
Prerequisites
Use a
¼
-in Allen screwdriver for all alignment screws.
Installation
Reference Manual
May 2022
00809-0100-4035
36
Emerson.com/Rosemount
Summary of Contents for Rosemount 935
Page 8: ...Product overview Reference Manual May 2022 00809 0100 4035 8 Emerson com Rosemount ...
Page 16: ...Technical description Reference Manual May 2022 00809 0100 4035 16 Emerson com Rosemount ...
Page 22: ...Operating modes Reference Manual May 2022 00809 0100 4035 22 Emerson com Rosemount ...
Page 46: ...Operation Reference Manual May 2022 00809 0100 4035 46 Emerson com Rosemount ...
Page 52: ...Troubleshooting Reference Manual May 2022 00809 0100 4035 52 Emerson com Rosemount ...
Page 56: ...Ordering information Reference Manual May 2022 00809 0100 4035 56 Emerson com Rosemount ...
Page 62: ...Specifications Reference Manual May 2022 00809 0100 4035 62 Emerson com Rosemount ...
Page 70: ...Wiring configurations Reference Manual May 2022 00809 0100 4035 70 Emerson com Rosemount ...
Page 74: ...Support Reference Manual May 2022 00809 0100 4035 74 Emerson com Rosemount ...