7.7
NQ Ratio below the permitted limit
Host status = F
0-20 = 1 mA
Light-emitting diode (LED) = yellow, blinking
Table 7-7: Potential cause and resolution
Potential cause
Resolution
Gas in path
Ensure that the path is clean and the weather
conditions are good.
7.8
NQ Ratio above the permitted limit
Host status = F
0-20 = 1 mA
LED = yellow, blinking
Table 7-8: Potential cause and resolution
Potential cause
Resolution
Poor alignment
Align.
7.9
Ratio out of the limit
Host status = F
0-20 = 1 mA
LED = yellow, blinking
Table 7-9: Potential causes and resolutions
Potential cause
Resolution
Poor alignment
Align.
Dirt on the window
Clean the window.
Detector fault
Replace or repair the detector.
7.10
Source fault
Yellow light-emitting diode (LED) blinking at the source
Table 7-10: Potential Causes and Resolutions
Potential cause
Resolution
Low/high input voltage
Check the power supply and installation.
Internal fault
Replace the source.
Reference Manual
Troubleshooting
00809-0100-4035
May 2022
Rosemount 935
51
Summary of Contents for Rosemount 935
Page 8: ...Product overview Reference Manual May 2022 00809 0100 4035 8 Emerson com Rosemount ...
Page 16: ...Technical description Reference Manual May 2022 00809 0100 4035 16 Emerson com Rosemount ...
Page 22: ...Operating modes Reference Manual May 2022 00809 0100 4035 22 Emerson com Rosemount ...
Page 46: ...Operation Reference Manual May 2022 00809 0100 4035 46 Emerson com Rosemount ...
Page 52: ...Troubleshooting Reference Manual May 2022 00809 0100 4035 52 Emerson com Rosemount ...
Page 56: ...Ordering information Reference Manual May 2022 00809 0100 4035 56 Emerson com Rosemount ...
Page 62: ...Specifications Reference Manual May 2022 00809 0100 4035 62 Emerson com Rosemount ...
Page 70: ...Wiring configurations Reference Manual May 2022 00809 0100 4035 70 Emerson com Rosemount ...
Page 74: ...Support Reference Manual May 2022 00809 0100 4035 74 Emerson com Rosemount ...