5.4.1
Signal limitation values
Table 5-1: Maintenance Channels' Limits
Channel
Installation distance
Minimum
Medium
Maximum
Reference
1 V gain 1
1 V gain 2
1 V gain 5
Signal
1 V gain 1
1 V gain 2
1 V gain 5
Ratio
0.6 - 1.4
0.6 - 1.4
0.6 - 1.4
NQRat
0.98 - 1.02
Lower explosive limit (LEL)
0 LEL x m
Temperature
Up to 25 °C beyond ambient temperature
Voltage
32 Vdc > V > 18 Vdc
Note
The installation information refers to the installation distance.
Minimum
The minimum distance, as defined according to the model number.
Medium
Half the maximum distance, as defined according to the model number.
Maximum
The maximum distance, as defined according to the model number.
Reference Manual
Operation
00809-0100-4035
May 2022
Rosemount 935
41
Summary of Contents for Rosemount 935
Page 8: ...Product overview Reference Manual May 2022 00809 0100 4035 8 Emerson com Rosemount ...
Page 16: ...Technical description Reference Manual May 2022 00809 0100 4035 16 Emerson com Rosemount ...
Page 22: ...Operating modes Reference Manual May 2022 00809 0100 4035 22 Emerson com Rosemount ...
Page 46: ...Operation Reference Manual May 2022 00809 0100 4035 46 Emerson com Rosemount ...
Page 52: ...Troubleshooting Reference Manual May 2022 00809 0100 4035 52 Emerson com Rosemount ...
Page 56: ...Ordering information Reference Manual May 2022 00809 0100 4035 56 Emerson com Rosemount ...
Page 62: ...Specifications Reference Manual May 2022 00809 0100 4035 62 Emerson com Rosemount ...
Page 70: ...Wiring configurations Reference Manual May 2022 00809 0100 4035 70 Emerson com Rosemount ...
Page 74: ...Support Reference Manual May 2022 00809 0100 4035 74 Emerson com Rosemount ...