4.10
Flash source wiring
4.10.1
Install flash source wiring
To install the wiring:
Figure 4-6: Source with Cover Removed
A. Housing
B. Internal earth connection
C. Terminal board
D. Inlet conduit
E. Detector holding plate
F. Earth terminal
Procedure
1. Release the back screw bolt and open the source back cover.
2. Remove the protective plug mounted on the source conduit/cable entry inlet; pull
the wires through the source inlet.
3. Use a
¾
-in.-14 national pipe thread (NPT) or M25 x 1.5 explosion-proof conduit
connection/cable gland to assemble the cable/explosion-proof conduit to the
detector.
4. Connect the wires to the required terminals according to the wiring diagram.
See
.
5. Connect the grounding wire to the ground screw outside the detector.
The source must be well grounded to earth ground.
Reference Manual
Installation
00809-0100-4035
May 2022
Rosemount 935
35
Summary of Contents for Rosemount 935
Page 8: ...Product overview Reference Manual May 2022 00809 0100 4035 8 Emerson com Rosemount ...
Page 16: ...Technical description Reference Manual May 2022 00809 0100 4035 16 Emerson com Rosemount ...
Page 22: ...Operating modes Reference Manual May 2022 00809 0100 4035 22 Emerson com Rosemount ...
Page 46: ...Operation Reference Manual May 2022 00809 0100 4035 46 Emerson com Rosemount ...
Page 52: ...Troubleshooting Reference Manual May 2022 00809 0100 4035 52 Emerson com Rosemount ...
Page 56: ...Ordering information Reference Manual May 2022 00809 0100 4035 56 Emerson com Rosemount ...
Page 62: ...Specifications Reference Manual May 2022 00809 0100 4035 62 Emerson com Rosemount ...
Page 70: ...Wiring configurations Reference Manual May 2022 00809 0100 4035 70 Emerson com Rosemount ...
Page 74: ...Support Reference Manual May 2022 00809 0100 4035 74 Emerson com Rosemount ...