Quick Installation Guide
00825-0100-4004, Rev DB
December 2011
Rosemount 8800D
6
NOTE
The required bolt load for sealing the gasket joint is affected by several factors, including
operating pressure, gasket material, thickness, and condition. A number of factors also
affect the actual bolt load resulting from a measured torque, including condition of bolt
threads, friction between the nut head and the flange, and parallelism of the flanges. Due to
these application-dependent factors, the required torque for each application may be
different. Follow the guidelines outlined in the ASME PCC-1 for proper bolt tightening. Make
sure the flowmeter is centered between flanges of the same nominal size as the flowmeter.
Insert Integral Temperature Sensor (MTA Option Only)
The Temperature Sensor is coiled and attached to the electronics bracket. Remove the
plastic band holding the sensor to the electronics bracket and insert temperature sensor into
into the hole at the bottom of the meter body. There is no need to remove the opposite end
from the electronics. Tighten with
1
/
2
inch open-end wrench approximately 1
1
/
4
turns past
finger-tight.
Remote Electronics
If you order one of the remote electronics options (options R10, R20, R30, R33, R50, or
RXX), the flowmeter assembly ships in two parts:
1. The meter body with an adapter installed in the support tube and an interconnecting
coaxial cable attached to it.
2. The electronics housing installed on a mounting bracket.
Mounting
Mount the meter body in the process flow line as described earlier in this section. Mount the
bracket and electronics housing in the desired location. The housing can be repositioned on
the bracket to facilitate field wiring and conduit routing.
Cable Connections
Refer to Figure 8 and the instructions on page 7 to connect the loose end of the coaxial
cable to the electronics housing.
Figure 8. Remote Electronics Installation
NOTE
Consult factory for SST installation
1
/2 NPT Conduit Adapter or Cable
Gland (Supplied by Customer)
Coaxial Cable
Meter Adapter
Union
Washer
Nut
Sensor Cable Nut
Support
Tube
Meter Body
Electronics
Housing
Ground Connection
Housing Base
Screw
Housing
Adapter
Housing Adapter
Screws
Conduit Adapter
(optional - supplied
by customer)
Coaxial
Cable Nut
4004_RevDB.fm Page 6 Wednesday, November 23, 2011 3:06 PM