Quick Installation Guide
00825-0100-4004, Rev DB
December 2011
Rosemount 8800D
8
CAUTION
Do not rotate the housing while the sensor cable is attached to the base of the housing. This
will stress the cable and may damage the sensor.
6. Place the electronics housing into the top of the support tube.
7. Use a 5/32” hex wrench to turn the three housing rotation screws counter-clockwise
(outward) to engage the support tube.
S
TEP
3: S
ET
J
UMPERS
Adjust jumpers to desired settings.
HART
If alarm and security jumpers are not installed, the flowmeter will operate normally with the
default alarm condition alarm
high
and the security
off
.
Figure 9. HART Jumpers and LCD
F
OUNDATION
fieldbus
If security and simulate enable jumpers are not installed, the flowmeter will operate normally
with the default security
off
and simulate enable
off
.
Figure 10. F
OUNDATION
fieldbus Jumpers and LCD
ALARM
HI
LO
ON
OFF
SECURITY
SIMULATE ENABLE
OFF
ON
ON
OFF
SECURITY