81
Rosemount 8600 Vortex Flow Meter
Reference Manual
May 2019
Troubleshooting
00809-0100-4860, Rev BF
5.6.5
Remote electronics procedure
If the Rosemount 8600D electronics housing is mounted remotely, some replacement
procedures are different than for the flowmeter with integral electronics. The following
procedures are exactly the same:
Replacing the Terminal Block in the Housing (see
).
Replacing the Electronics Boards (see
).
Replacing the Sensor (see
).
To disconnect the coaxial cable from the meter body and electronics housing, follow the
instructions below.
Disconnect the coaxial cable at the meter
1.
Remove the access cover on the meter body bracket if present.
2.
Loosen the three housing rotation screws at the base of the meter adapter with a
5
/
32
-in. hex wrench by turning the screws clockwise (inward) until they clear the
bracket.
3.
Slowly pull the meter adapter no more than 1.5-in. (40 mm) from the top of the
bracket.
4.
Loosen and disconnect the sensor cable nut from the union using
a
5
/
16
-in. open end wrench.
Note
Do not pull the adapter more than 1.5-in. (40 mm) from the top of the bracket. Damage to
the sensor may occur if the sensor cable is stressed.