Reference Manual
00809-0100-4860, Rev BF
Electronics Verification
May 2019
116
Rosemount 8600 Vortex Flow Meter
Example 3 (SI units)
Therefore, an input frequency of 700.00 Hz represents a flow rate of 9259.2 NCMH in this
application.
For a given input frequency, you may also determine the current output. Use the prior
example with an input frequency of 375 Hz.
URV= 10000 NCMH LRV= 0 NCMH F
in
(Hz) = 375.00
= 11.94 mA
Therefore, an input frequency of 375.00 Hz represents a current output of 11.94 mA.
Fluid
= Natural gas
URV
= 10,000 NCMH
Line size
= 80 mm
LRV
= 0 NCMH
Line pressure
= 1000 KPas
C
20
= C
x
/sp. units factor
)
Operating
Temperature
= 10 °F
Density (
)
= 9.07754 kg/cu-mtr
(operating)
Viscosity
= 0.01 cp
Input
Frequency
= 700 Hz
K-factor
(compensated)
= 10.797 (via Field
Communicator or AMS)
Q
= F(Hz) /(K x C
20
) where: C
20
= C
11
/10.48 (Density Ratio)
= 700/{10.797 x (.0733811/10.48)}
= 9259.2 NCMH
I
F Hz
K C
20
LRV
–
URV LRV
–
----------------------------------------------------------------
16
4
+
=
I
375
10.797 0.0070020
0
–
10000 0
–
--------------------------------------------------------------------------------------
16
4
+
=