Configuration
00809-0100-4860, Rev BF
Process temperature
Allows users to view the process temperature value when the vortex transmitter has the
temperature sensor option. Also allows the user to configure the process temperature
units.
Process temperature
Displays the current process temperature value.
Process temperature units
Allows the user to configure the units for the process temperature from the available list.
deg C
deg F
deg R
Kelvin
T/C failure mode
Allows the user to configure the temperature sensor failure mode. In the event that the
thermocouple sensor fails, the vortex can go either into an alarm output mode, or continue
to operate normally using the Fixed Process Temperature value. See Fixed Process
Temperature
. This mode is only relevant with the MTA option.
Note
If the Primary Variable is set to Process Temperature and there is an error, the output will
always go to alarm and this setting will be ignored.
Cold Junction (CJ) temperature
Allows users to view the thermocouple cold junction temperature value when the vortex
has the temperature sensor option. Also allows the user to configure the CJ temperature
units.
CJ temperature
Displays the current thermocouple cold junction temperature value.
FastKeys
1, 1, 4, 9
FastKeys
1, 1, 4, 9, 1
FastKeys
1, 1, 4, 9, 2
FastKeys
1, 1, 4, 9, 3
FastKeys
1, 1, 4, Scroll to bottom of list
FastKeys
1, 1, 4, -, 1