Attach the SuperModule
to the Housing
IMPORTANT
The V-Seal must be installed at the bottom of the housing.
1. Apply a light coat of low temperature silicon grease to the SuperModule
threads and o-ring.
2. Thread the housing completely onto the SuperModule. The housing must
be no more than one full turn from flush with the SuperModule to comply
with explosion-proof requirements.
3. Tighten the housing rotation set screw using a
3
/
32
-inch hex wrench.
Install Assembly in the
PlantWeb Housing
1. Apply a light coat of low temperature silicon grease to the SuperModule
connector.
2. Insert the SuperModule connector into the top of the SuperModule.
3. Gently slide the assembly into the housing, making sure the pins from
the PlantWeb housing properly engage the receptacles on the assembly.
4. Tighten the captive mounting screws.
5. Attach the PlantWeb housing cover and tighten so that metal contacts
metal to meet explosion-proof requirements.
Install the Terminal Block
PlantWeb Housing
1. Gently slide the terminal block into the housing, making sure the pins
from the PlantWeb housing properly engage the receptacles on the
terminal block.
2. Tighten the captive screws on the terminal block.
3. Attach the PlantWeb housing cover and tighten so that metal contacts
metal to meet explosion-proof requirements.
Device ID
Locate a device I.D. for a known tag:
1. Log in to iTrax
®
2. Click Device on the left hand side under the Configuration Menu
3. Choose the device from the Select tag to configure menu
Allow iTraX
®
to load the device configuration options
4. Click Measurement
5. The device I.D. appears in the list of Measurement Configuration options
Summary of Contents for Rosemount 753R
Page 2: ......
Page 4: ......
Page 10: ...Reference Manual 00809 0100 4379 Rev AA March 2006 Rosemount 753R Indicator 1 4 ...
Page 20: ...Reference Manual 00809 0100 4379 Rev AA March 2006 Rosemount 753R Indicator 2 10 ...
Page 27: ...Reference Manual 00809 0100 4379 Rev AA March 2006 3 7 Rosemount 753R Indicator ...
Page 28: ...Reference Manual 00809 0100 4379 Rev AA March 2006 Rosemount 753R Indicator 3 8 ...
Page 43: ...Reference Manual 00809 0100 4379 Rev AA March 2006 5 7 Rosemount 753R Indicator ...
Page 44: ...Reference Manual 00809 0100 4379 Rev AA March 2006 Rosemount 753R Indicator 5 8 ...
Page 50: ...Reference Manual 00809 0100 4379 Rev AA March 2006 Rosemount 753R Indicator 6 6 ...
Page 56: ...Reference Manual 00809 0100 4379 Rev AA March 2006 Rosemount 753R Indicator A 6 ...
Page 58: ...Reference Manual 00809 0100 4379 Rev AA March 2006 Rosemount 753R Indicator B 2 ...
Page 59: ......