5
Reference Manual
00809-0100-4728, Rev MB
Introduction
July 2016
Introduction
Note
If a hazardous substance is identified, a Material Safety Data Sheet (MSDS), required by law to be
available to people exposed to specific hazardous substances, must be included with the returned
materials.
1.5
Transmitter security
1.5.1
Available security options
There are three security methods to utilize with the Rosemount 644 Transmitter.
Software Security Switch (Write Protect)
HART Lock
LOI Password
The
Write Protect
feature allows you to protect the transmitter data from accidental or unwarranted
configuration changes. To enable the write protect feature, perform the following procedures.
Configuring Write Protect, HART Lock and LOI Password with a Field
Communicator
Configuring Write Protect, HART Lock and LOI Password with AMS
™
Device Manager
Right click on the device and select the
Configure
menu.
1.
In the left navigation pane choose
Manual Setup
then chose the
Security tab
.
2.
All three parameters can be configured from this screen.
3.
Click
Apply
when complete.
From the
HOME
screen, enter the Fast Key sequence
Device Dashboard Fast Keys-Write Protect
2, 2, 9, 1
Device Dashboard Fast Keys-HART Lock
2, 2, 9, 2
Device Dashboard Fast Keys- LOI Password
2, 2, 9, 3
Summary of Contents for Rosemount 644 HART
Page 2: ......
Page 16: ...6 Reference Manual 00809 0100 4728 Rev MB Introduction July 2016 Introduction ...
Page 86: ...76 Reference Manual 00809 0100 4728 Rev MB Troubleshooting July 2016 Troubleshooting ...
Page 138: ...128 Reference Manual 00809 0100 4728 Rev MB Index July 2016 Index ...
Page 139: ......