101
Product Certifications
Product Certifications
July 2016
Reference Manual
00809-0100-4728, Rev MB
Appendix B
Product Certifications
Approved manufacturing locations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 101
European Union Directive Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 101
Rosemount 644 with Foundation Fieldbus and Profibus PA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 103
B.1 Approved manufacturing locations
Emerson Process Management Rosemount Division. –
Chanhassen, MN
Emerson Process Management Asia Pacific Limited – Singapore
Emerson Process Management GmbH & Co. - Karlstein, Germany
Emerson Process Management (India) Private Ltd. – Mumbai, India
Emerson Process Management Brazil - Sorocaba, Brazil
Emerson Process Management, Dubai - Emerson FZE
Beijing Rosemount Far East Instrument Co., Limited. - Beijing,
China
B.2 European Union Directive
Information
B.2.1 Ordinary Location Certification from FM
Approvals
As standard, the transmitter has been examined and tested to
determine that the design meets basic electrical, mechanical, and
fire protection requirements by FM Approvals, a nationally
recognized testing laboratory (NRTL) as accredited by the Federal
Occupational Safety and Health Administration (OSHA).
B.2.2 Hazardous Locations Certifications
North American Certifications
FM Approvals
I5
Intrinsically Safe and Non-Incendive
Certificate: 3044581
Standards: FM Class 3600 – 1998, FM Class 3610 – 2010, FM
Class 3611 – 2004,
FM Class 3810 – 2005, ANSI/NEMA
®
250 - 2003
Markings: (without enclosure):
INT. SAFE CL I, GP ABCD, T4
IS CL I Zone 0, AEX ia IIC; T4 Ga
NI CL I, DIV.2, GP ABCD
INSTALL PER DRAWINGS 00644-2071
Markings: (with enclosure):
IS CL I,II,III, GP ABCDEFG T4
IS CL I Zone 0, AEX ia IIC; T4
NI CL I, DIV.2, GP ABCD
INSTALL PER DRAWINGS 00644-2071
ENCLOSURE TYPE 4X
Special Conditions for Safe Use (X):
1. The surface resistivity of the non-metallic enclosure
materials is greater than one gigaohm. Care must be taken to
avoid electrostatic charge build-up. Do not rub or cleaned
with solvents or a dry cloth.
2. The Model 644 optional enclosures may contain aluminum
and are considered a potential risk of ignition by impact or
friction. Care must be taken into account during installation
and use to prevent impact and friction.
E5
Explosion-Proof and Dust Ignition Proof
Certificate: 3006278
Standards: FM Class 3600 – 1998, FM Class 3615 – 2006, FM
Class 3810 – 2005, ANSI/NEMA 250 - 2003
Markings:EXPLOSIONPROOF FOR CL. I, DIV. 1, GP BCD
DUST-IGNITIONPROOF FOR CL. II & III, DIV. 1, GP
EFG
NON-INCENDIVE FOR CL. I, DIV 2, GP ABCD
WHEN INSTALLED PER ROSEMOUNT DRAWING
00644-1049
CONDUIT SEAL NOT REQUIRED;
ENCLOSURE TYPE 4X
CSA International
I6
Intrinsically
Safe
Certificate: 1091070
Standards: CSA Std. C22.2 No. 142 – M1987, CSA Std. C22.2
No. 157 – 92, ANSI/ISA 12.27.02-2003
Markings: (without enclosure):
Ex ia
CLASS I (GRPS A, B, C, D) CLASS I, ZONE 0, IIC
SUITABLE FOR CLASS I DIV 2, GROUPS A, B, C, D
INSTALL PER DRAWING 00644-2072.
Summary of Contents for Rosemount 644 HART
Page 2: ......
Page 16: ...6 Reference Manual 00809 0100 4728 Rev MB Introduction July 2016 Introduction ...
Page 86: ...76 Reference Manual 00809 0100 4728 Rev MB Troubleshooting July 2016 Troubleshooting ...
Page 138: ...128 Reference Manual 00809 0100 4728 Rev MB Index July 2016 Index ...
Page 139: ......