Reference Manual
00809-0100-4716, Rev JA
May 2008
2-17
Rosemount 3095 MultiVariable
Install RTD Assembly
(optional)
4.
(Optional) Install the Series 68 or Series 78 RTD Assembly.
NOTE
To meet ISSep/CENELEC Flameproof certification, only European
Flameproof Cable Assemblies (Process Temperature Input Codes A, B, or C)
may be used for RTD cable installation.
a. Mount the RTD Assembly in the desired location. Refer to the
appropriate primary element standard concerning recommended
RTD installation location.
b. Connect the RTD cable (optional) to the 3095 RTD connector. All
RTD 3095 Cable assemblies use the 3095 RTD cable connector.
Identify the cable type being installed and follow the steps below.
Figure 2-8. Armored Shielded
RTD Cable Assembly
First, fully engage the
black cable connector to
the 3095 RTD connector
pins.
Second, screw in and
tighten the cable adapter
until metal to metal
contact occurs. Install
compression fitting.
Third, use pliers to screw
in and tighten the strain
relief cap onto the
compression fitting.
3
/
4
to
1
/
2
-14
NPT Adapter
(screws into RTD
Connection Head)
Washer
Black Cable Connector
(connect to 3095 RTD Pins)
Compression
Fitting
Compression Fitting
Cable Adapter
Bushing
Cap
Conductive Bushing
(slide stop to edge of
armored cable)
RTD Cap
Cable Wires
(2) Red wires
(2) White wires
Summary of Contents for Rosemount 3095
Page 2: ......
Page 8: ......
Page 130: ...Reference Manual 00809 0100 4716 Rev JA May 2008 Rosemount 3095 MultiVariable 5 24 ...