Related information
5.5 Review data for all calibrations
ProLink III
Device Tools
→
Calibration Data
Field communicator
Configure
→
Manual Setup
→
Calibration Factors
For all calibrations, you can review the gas data and sensor time period data that was used to calculate the
calibration factors, and you can review the calibration factors.
5.6 Change the label for the active calibration
ProLink III
Device Tools
→
Calibration Data
Field communicator
Configure
→
Manual Setup
→
Calibration Factors
You can change the default calibration labels.
Procedure
1. Select and delete the value in the
Calibration
x
Label
field.
2. Enter your preferred calibration label and press
Apply
.
5.7 Select the active calibration
ProLink III
Device Tools
→
Calibration Data
→
Active Calibration
Field communicator
Configure
→
Manual Setup
→
Measurements
→
Gas Calibration
→
Active Calibration
The SGM can store up to four calibrations. The active calibration specifies the calibration to be used for
measurement.
Important
Do not confuse the active calibration with the calibration being performed. For example, you may perform
Calibration 4, and then choose to use Calibration 3 for measurement.
Procedure
Set
Active Calibration
to the calibration you want to use for measurement.
Related information
Review data for all calibrations
Purging and calibration
Configuration and Use Manual
March 2021
MMI-20020954
44
Micro Motion Gas Specific Gravity Meters (SGM)