c) Set the calibration gas selector valve (A) to the calibration gas 2 inlet (I) position. The process
gas inlet valve (D) must remain closed.
d) Activate
SCROLL
until
ENTER GAS HIGH
appears, then activate
Select
.
e) Enter the specific gravity, molecular weight, or relative density of the calibration gas that is
flowing through the system and store the value by activating both
SCROLL
and
SELECT
at same
time until the previous menu appears.
f) Activate
SCROLL
g) When
CAL GAS HIGH
appears, activate
Select
to start the calibration.
h) During the calibration, activate
SCROLL
to observe the
Sensor Time Period
and
Stability
values
during the calibration.
i) Wait between 3 to 15 minutes for the system to stabilize. When
Stability
is
Good
, activate
Select
.
If measurement does not stabilize after 30 minutes, activate
SCROLL
to abort the calibration.
j) Activate
Select
again to accept the calibration value.
k) As the gas flows, select the
Start
button and observe the
Sensor Time Period
and
Stability
values.
l) Disconnect the lower calibration gas.
9. Activate
SCROLL
until
CALC K VAL
appears, then activate
Select
.
The meter automatically calculates the calibration factors from the stored data.
10. View the calibration factors.
a) When
RESULT DISPLAY
appears, activate
Select
.
b) Activate
SCROLL
to view the calibration factors and data.
Results are displayed in this order:
•
The specific gravity, molecular weight, or relative density used to calculate the K0 calibration
factor
•
The time period used to calculate the K0 calibration factor
•
The specific gravity, molecular weight, or relative density used to calculate the K1 calibration
factor (3-point calibrations only)
•
The time period used to calculate the K1 calibration factor (3-point calibrations only)
•
The specific gravity, molecular weight, or relative density used to calculate the K2 calibration
factor
•
The time period used to calculate the K2 calibration factor
•
The K0 calibration factor
•
The K1 calibration factor (3-point calibrations only), displayed in exponential format
•
The K2 calibration factor, displayed in exponential format
Purging and calibration
Configuration and Use Manual
March 2021
MMI-20020954
36
Micro Motion Gas Specific Gravity Meters (SGM)