Maintenance and Troubleshooting Manual
Section 3: Meter maintenance
3-9000-772 Rev B
January 2018
Replacing the Acquisition Module
69
13.
Reattach the Bracket/Cover to the Base Enclosure with the four hex head bolts and lock
washers. Tighten bolts with a 7/16 inch (11mm) wrench.
14.
Reapply power and connect to the meter with Daniel MeterLink. Open the Meter
Monitor (Detailed) view and verify the meter is acquiring data, the transducers have
good signals and the chords flow profiles are good.
15.
Run the Zero Flow Calibration Wizard; select the
Calibrate|Zero Calibrate
menu and
generate the report. Then, collect and save a Maintenance log to show the “as left”
condition of the meter.
Or use AMS
TM
Device Manager or a HART® Field Communicator and select
Overview
|Zero Calibration|Start
menu or
Service Tools|Maintenance| Zero Calibration
|
Start.
16.
Close your application connection (Daniel MeterLink, AMS
TM
Device Manager or HART®
Field Communicator).
17.
Install the end cap and the end cap security latch using a 3 mm Allen wrench.
18.
If required, install the security seals (
see Section 3.6.2
) through the socket head cap
screw in the Bracket/Cover.
19.
This completes the Acquisition Module replacement procedure.