Maintenance and Troubleshooting Manual
Section 3: Meter maintenance
3-9000-772 Rev B
January 2018
Re-wetting the transducers
47
3.
Use a 7/16 inch wrench to loosen the lock nut on the set screw. Hold the set screw
using a 1/8 inch Allen wrench as the lock nut is loosened a few turns. Use a 1/8 inch to
loosen, but do not remove, the half dog set screw. See the figure below and refer to the
LT-07 Transducer Assembly
Figure 3-7
.
Figure 3-9 Transducer assembly removal
4.
Mark the cable position on the meter body. This mark will be helpful to rotationally
align the transducer with the transducer housing during re-assembly.
5.
Pull the transducer assembly out of the housing.
Figure 3-10 LT-07 Transducer assembly removed from transducer housing
A. Half dog set screw and hex nut
B. Transducer cover connection head
B.
A.
A.
A. Transducer assembly