Reference Manual
00809-0100-4793, Rev CA
August 2004
5-3
Rosemount 8742C
1.
Tube Calibration Number - 1000015010000000
2.
Units - ft/s
3.
Coil Pulse Mode - 5 Hz
The instructions for changing these parameters are located in the parameter
descriptions in this section.
Set the loop to
manual
, if necessary, before you begin. Complete the following
steps:
1.
Power down the transmitter.
2.
Connect the transmitter to a Rosemount 8714D flowtube simulator.
3.
Power up the transmitter with the Rosemount 8714D connected and
read the flow rate. The electronics need about a 30-minute warm-up
time to stabilize.
4.
Initiate an electronics trim through the transducer block. Click on the
transducer block, select calibrate, and then select Electronics Trim.
The electronics trim takes about two minutes to complete. No transmitter
adjustments are required. A symbol appears in the lower right-hand corner of
the local display to indicate the process is running.