Reference Manual
00809-0100-4793, Rev CA
August 2004
3-13
Rosemount 8742C
NOTE
Consult factory for installations requiring cathodic protection or situations
where there are high currents or high potential in the process.
The flowtube case should always be earth grounded in accordance with
national and local electrical codes. Failure to do so may impair the protection
provided by the equipment. The most effective grounding method is direct
connection from the flowtube to earth ground with minimal impedance.
The Internal Ground Connection (Protective Ground Connection) located in
side the junction box is the Internal Ground Connection screw. This screw is
identified by the ground symbol:
Figure 3-13. No Grounding
Options or Grounding Electrode
in Lined Pipe
Table 3-6. Grounding Installation
Grounding Options
Type of Pipe
No Grounding Options
Grounding Rings
Grounding Electrodes
Lining Protectors
Conductive Unlined Pipe
Not Required
Not Required
Conductive Lined Pipe
Insufficient Grounding
Non-Conductive Pipe
Insufficient Grounding
See Figure 3-15
Earth
Ground
8705-0040C