Decommissioning
52
12 Decommissioning
The Autoloader is decommissioned as follows:
1. Ensure that the Autoloader has been switched off (see Pos. (
2
) in
“
Operating side
”).
2. Pull out the plug (see Pos. (
2
) in
”).
3. Pull the compressed air hose off (see Pos. (
1
) in
”).
4. Dismantle the compressed air hoses to the analyzer.
5. Pull the cable for the interface to the analyzer off (see Pos. (
3
) in
”).
6. Empty the crucible bin.
7. Place the empty crucibles on the crucible tray.
This ensures that the crucibles do not get lost and are available next time the Autoloader is
commissioned.
The Autoloader has been decommissioned.
13 Storage
The following storage conditions apply to storage of the Autoloader CS-i/CS-d:
•
Indoor area
•
Ambient temperature b5 and +40 °C
•
Ambient humidity <50 % at +40 °C, non-condensing
Store the Autoloader as follows:
1. Decommission the Autoloader, see
”.
2. Store the Autoloader under the specified storage conditions.
The Autoloader has been stored.