5.4 Replacing the side wall of the analyzer
DANGER
D3.0000
Risk to life caused by an electric shock
Exposed power contacts
−
Contact with exposed power contacts can cause life-threatening injuries
or death.
•
Turn the main switch to Position 0 and pull out the power plug prior
to maintenance work on the analyzer.
In order to enable the Autoloader to be screwed to the contour of the analyzer profile, the side
wall of the analyzer must first be replaced by the supplied side wall.
Proceed as follows to do this.
Fig. 14: Side wall with and without cut-outs
Position
Component
Description
1
Side wall screws (2x)
For securing the side wall of the analyzer
2
Side wall without cut-outs
Side wall supplied with the CS-i/CS-d
3
Side wall with cut-outs
Side wall supplied with the Autoloader
1. Remove the screws (
1
). The side wall continues to be held by two brackets at the bottom.
2. Lift the side wall (
2
) up a few centimetres to lift the retaining bracket out of the profile, then
place it to one side. This side wall will no longer be needed.
3. Hang the prepared side wall with cut-outs supplied for the Autoloader (
3
) into the profile and
secure it using the screws (
1
).
The side wall has been replaced.