10 Cleaning
WARNING
W4.0003
Risk to life caused by an electric shock
Using water to clean live parts
•
Using water to clean the Autoloader can lead to life-threatening injuries
caused by an electric shock if the Autoloader has not been
disconnected from the power supply.
•
Only clean the Autoloader with water when the Autoloader has been
disconnected from the power supply.
The Autoloader CS-i/CS-d must be inspected regularly for spilt sample material.
Remove spilt sample material as follows:
1. Ensure that the Autoloader has been switched off (see Pos. (
2
) in
“
Operating side
”).
2. Remove the spilt sample material using an appropriate tool for the properties of the
material, e.g. tweezers, vacuum cleaner or damp cloth.
3. Wait until the Autoloader is completely dry if necessary.
4. Switch the Autoloader back on (see Pos. (
2
) in
“
Operating side
”).
The spilt material has been removed.