Safety
14
CAUTION
C5.0000
Risk of poisoning and suffocation
Toxic substances
−
Hazardous liquids, gases, mist, vapours or dust may be present when
working on the Autoloader; these may cause poisoning or suffocation if
there is contact with them or they are inhaled.
•
Where necessary, conduct a risk assessment for the samples.
•
Where necessary, wear suitable goggles or protective gloves.
•
Ensure there is sufficient ventilation.
•
Never eat or drink near toxic substances.
CAUTION
C6.0000
Risk of injury
Flammable samples
−
Flammable or explosive samples may explode and cause injuries.
−
Where necessary, conduct a risk assessment for the samples.
−
Where necessary, wear suitable goggles or protective gloves.
−
Ensure there is sufficient ventilation.
CAUTION
C7.0000
Risk of burns
Flammable surface
−
Crucibles can fall onto the surface or onto the floor, e.g. if the compressed
air fails or if the crucible bin is missing. A fire may result if there is a
flammable surface beneath the Autoloader or on the floor, causing
injuries.
•
Place the Autoloader on a fireproof surface.
•
Use fireproof flooring.
•
Operation of the Autoloader should always be supervised.
CAUTION
C8.0000
Risk of posture problems
Lifting the Autoloader unaided
−
The weight of the Autoloader can cause injuries resulting from posture
problems due to an unhealthy posture or inattentiveness when lifting.
•
Ensure you maintain a healthy posture when lifting the Autoloader.