Installation
30
5.5 Installing the furnace sensor
The supplied magnetic switch must be attached to the pneumatic cylinder of the furnace so that
the autoloader can recognize the state of the furnace (open / closed).
Proceed as follows to do this.
Fig. 15: Furnace Sensor
Position
Component
Description
1
Lower Furnace Cover
2
Lower Furnace Cover Screws (4)
3
Furnace Sensor with Bracket
Furnace Sensor w/ Bracket supplied with the
Autoloader
4. Ensure that the furnace of the analyzer is open.
5. Remove the screws (
2
) and the lower furnace cover (
1
).
6. Attach the magnetic sensor to the fixing bracket. Both parts are supplied with the
autoloader.
7. Place sensor and the bracket (
3
) at the pneumatic cylinder as shown above.
8. Adjust the correct height by moving the sensor vertical up and down. The sensor must
switch (closed contact) while the furnace is fully open.
Remark: The sensor must switch (open contact) while the furnace is closed or partial
closed.
9. Secure the sensor with the screw of the bracket at the working position.
10. Place the lower furnace cover (
1
) at the analyzer and secure it using the screws (
2
).
The furnace sensor has been installed.
2
1
3