AM852 Veterinary Anaesthesia machine User’ Manual
20
3. If more than one the evaporators can be opened at the same time, these
evaporators should be removed and reinstalled according to the operation
manual of evaporators.
4.6Test Respiration system
1. Verify that the anaesthetic gas spiritualization system(AGSS) have been
connected correctly and work normally.
2.
Expiration unilateral valve 1 ascends during the inspiration period and
descends during the expiration period.
Inspiration unilateral valve 2 ascends during the inspiration period and
descends during the expiration period.
Figure 4-1 Respiration System
3. Gas leak test of gas bag circuit:
·
Wrench manual/ mechanical ventilator control to manual ventilation.
·
Set the flowrate of O2 to about 0.3L/min and close others flowrate
regulating switches.
·
Close APL valve.
·
Install manual gas bag on the port of it.
·
Connect the Y-junction with the simulated lung at the animal interface.
·
Press the O
2
flush button until the pressure gauge reaches to about 3kPa.